Title: SU-E-T-305: Study of the Eclipse Electron Monte Carlo Algorithm for Patient Specific MU Calculations

Abstract

Purpose: To evaluate the Eclipse electron Monte Carlo (eMC) algorithm based on patient specific monitor unit (MU) calculations, and to propose a new factor which quantitatively predicts the discrepancy of MUs between the eMC algorithm and hand calculations. Methods: Electron treatments were planned for 61 patients on Eclipse (Version 10.0) using the eMC algorithm for Varian TrueBeam linear accelerators. For each patient, the same treatment beam angle was kept for a point dose calculation at dmax performed with the reference condition, which used an open beam with a 15×15 cm2 size cone and 100 SSD. A patient specific correction factor (PCF) was obtained by getting the ratio between this point dose and the calibration dose, which is 1 cGy per MU delivered at dmax. The hand calculation results were corrected by the PCFs and compared with MUs from the treatment plans. Results: The MU from the treatment plans were in average (7.1±6.1)% higher than the hand calculations. The average MU difference between the corrected hand calculations and the eMC treatment plans was (0.07±3.48)%. A correlation coefficient of 0.8 was found between (1-PCF) and the percentage difference between the treatment plan and hand calculations. Most outliers were treatment plans with smallmore » beam opening (< 4 cm) and low energy beams (6 and 9 MeV). Conclusion: For CT-based patient treatment plans, the eMC algorithm tends to generate a larger MU than hand calculations. Caution should be taken for eMC patient plans with small field sizes and low energy beams. We hypothesize that the PCF ratio reflects the influence of patient surface curvature and tissue inhomogeneity to patient specific percent depth dose (PDD) curve and MU calculations in eMC algorithm.« less

Authors:
; ; ;  [1]
  1. UCLA School of Medicine, Los Angeles, CA (United States)
Publication Date:
OSTI Identifier:
22355864
Resource Type:
Journal Article
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 41; Journal Issue: 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; ALGORITHMS; ANIMAL TISSUES; BEAMS; DEPTH DOSE DISTRIBUTIONS; ELECTRONS; LINEAR ACCELERATORS; MONTE CARLO METHOD; PATIENTS; RADIATION DOSES

Citation Formats

Wang, X, Qi, S, Agazaryan, N, and DeMarco, J. SU-E-T-305: Study of the Eclipse Electron Monte Carlo Algorithm for Patient Specific MU Calculations. United States: N. p., 2014. Web. doi:10.1118/1.4888637.
Wang, X, Qi, S, Agazaryan, N, & DeMarco, J. SU-E-T-305: Study of the Eclipse Electron Monte Carlo Algorithm for Patient Specific MU Calculations. United States. https://doi.org/10.1118/1.4888637
Wang, X, Qi, S, Agazaryan, N, and DeMarco, J. 2014. "SU-E-T-305: Study of the Eclipse Electron Monte Carlo Algorithm for Patient Specific MU Calculations". United States. https://doi.org/10.1118/1.4888637.
@article{osti_22355864,
title = {SU-E-T-305: Study of the Eclipse Electron Monte Carlo Algorithm for Patient Specific MU Calculations},
author = {Wang, X and Qi, S and Agazaryan, N and DeMarco, J},
abstractNote = {Purpose: To evaluate the Eclipse electron Monte Carlo (eMC) algorithm based on patient specific monitor unit (MU) calculations, and to propose a new factor which quantitatively predicts the discrepancy of MUs between the eMC algorithm and hand calculations. Methods: Electron treatments were planned for 61 patients on Eclipse (Version 10.0) using the eMC algorithm for Varian TrueBeam linear accelerators. For each patient, the same treatment beam angle was kept for a point dose calculation at dmax performed with the reference condition, which used an open beam with a 15×15 cm2 size cone and 100 SSD. A patient specific correction factor (PCF) was obtained by getting the ratio between this point dose and the calibration dose, which is 1 cGy per MU delivered at dmax. The hand calculation results were corrected by the PCFs and compared with MUs from the treatment plans. Results: The MU from the treatment plans were in average (7.1±6.1)% higher than the hand calculations. The average MU difference between the corrected hand calculations and the eMC treatment plans was (0.07±3.48)%. A correlation coefficient of 0.8 was found between (1-PCF) and the percentage difference between the treatment plan and hand calculations. Most outliers were treatment plans with small beam opening (< 4 cm) and low energy beams (6 and 9 MeV). Conclusion: For CT-based patient treatment plans, the eMC algorithm tends to generate a larger MU than hand calculations. Caution should be taken for eMC patient plans with small field sizes and low energy beams. We hypothesize that the PCF ratio reflects the influence of patient surface curvature and tissue inhomogeneity to patient specific percent depth dose (PDD) curve and MU calculations in eMC algorithm.},
doi = {10.1118/1.4888637},
url = {https://www.osti.gov/biblio/22355864}, journal = {Medical Physics},
issn = {0094-2405},
number = 6,
volume = 41,
place = {United States},
year = {Sun Jun 01 00:00:00 EDT 2014},
month = {Sun Jun 01 00:00:00 EDT 2014}
}