Title: SU-E-T-242: Monte Carlo Simulations Used to Test the Perturbation of a Reference Ion Chamber Prototype Used for Small Fields

Purpose: To study the perturbation due to the use of a novel Reference Ion Chamber designed to measure small field dosimetry (KermaX Plus C by IBA). Methods: Using the Phase-space files for TrueBeam photon beams available by Varian in IAEA-compliant format for 6 and 15 MV. Monte Carlo simulations were performed using BEAMnrc and DOSXYZnrc to investigate the perturbation introduced by a reference chamber into the PDDs and profiles measured in water tank. Field sizes ranging from 1×1, 2×2,3×3, 5×5 cm2 were simulated for both energies with and without a 0.5 mm foil of Aluminum which is equivalent to the attenuation equivalent of the reference chamber specifications in a water phantom of 30×30×30 cm3 and a pixel resolution of 2 mm. The PDDs, profiles, and gamma analysis of the simulations were performed as well as a energy spectrum analysis of the phase-space files generated during the simulation. Results: Examination of the energy spectrum analysis performed shown a very small increment of the energy spectrum at the build-up region but no difference is appreciated after dmax. The PDD, profiles and gamma analysis had shown a very good agreement among the simulations with and without the Al foil, with a gamma analysismore » with a criterion of 2% and 2mm resulting in 99.9% of the points passing this criterion. Conclusion: This work indicates the potential benefits of using the KermaX Plus C as reference chamber in the measurement of PDD and Profiles for small fields since the perturbation due to in the presence of the chamber the perturbation is minimal and the chamber can be considered transparent to the photon beam.« less
Authors:
; ; ;  [1]
  1. Midsouth Radation Physics, Little Rock, AR (United States)
Publication Date:
OSTI Identifier:
22351149
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 41; Journal Issue: 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
71 CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS; ALUMINIUM; COMPUTERIZED SIMULATION; ENERGY SPECTRA; MONTE CARLO METHOD; PERTURBATION THEORY; PHASE SPACE; PHOTON BEAMS