Title: SU-E-T-169: Evaluation of Oncentra TPS for Nasopharynx Brachy Using Patient Specific Voxel Phantom and EGSnrc

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888498· OSTI ID:22339917
; ;  [1]
  1. Shiraz University, Shiraz (Iran, Islamic Republic of)

Purpose: Nasopharnx carcinoma (NPC) treatment is being carried out using Ir-192 HDR seeds in Mehdieh Hospital in Hamadan, Iran. The Oncentra™ TPS is based on optimized TG-43 formalism which disregards heterogeneity in the treatment area. Due to abundant heterogeneity in head and neck, comparison of the Oncentra™ TPS dose evaluation and an accurate dose calculation method in NPC brachytherapy is the objective of this study. Methods: CT DICOMs of a patient with NPC obtained from Mehdieh Hospital used to create 3D voxel phantom with CTCREATE utility of EGSnrc code package. The voxel phantom together with Ir-192 HDR brachytherapy source were the input to DOSXYZnrc to calculate the 3D dose distribution. The sources were incorporate with type 6 source in DOSXYZnrc and their dwell times were taken into account in final dose calculations. Results: The direct comparison between isodoses as well as DVHs for the GTV, PTV and CTV obtained by Oncentra™ and EGSnrc Monte Carlo code are made. EGSnrc results are obtained using 5×10{sup 9} histories to reduce the statistical error below 1% in GTV and 5% in 5% dose areas. The standard ICRP700 cross section library is employed in DOSXYZnrc dose calculation. Conclusion: A direct relationship between increased dose differences and increased material density (hence heterogeneity) is observed when isodoses contours of the TPS and DOSXYZnrc are compared. Regarding the point dose calculations, the differences range from 1.2% in PTV to 5.6% for cavity region and 7.8% for bone regions. While Oncentra™ TPS overestimates the dose in cavities, it tends to underestimate dose depositions within bones.

OSTI ID:
22339917
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English