Title: SU-E-J-162: Quality Assurance Procedures for MR Guided Focused Ultrasound Treatment of Bone Metastasis

Abstract

Purpose: The purpose of this work is to develop and verify our quality assurance (QA) procedures to ensure the safety and efficacy of MR-guided focused ultrasound (MRgFUS) treatment of bone metastases. Methods: A practical QA program was developed. Monthly and daily QA (DQA) procedures were performed. The major QA items included the checks of the machine hardware, software and patient safety features. Briefly, these checks/tests include: 1) the cooling system reservoir and treatment table; 2) power to the treatment table; 3) the MR coil; 4) the transducer position with MRI; 5) image display on the treatment work station; 6) the effective focal spot in 3 directions using MR thermometry; and 7) all the safety devices including a sonication lamp, and the emergency stop-sonication switches. In order to avoid patient skin burn, it is important to remove gas bubbles in the interfaces between the treatment table and the gel pad, and the gel pad and patients skin during the patient setup. Our QA procedures have been verified and evaluated through patient treatments. Seven patients with scapula, humeral head, sacrum, ilium, pubic ramus and acetabular bone metastases were treated using MRgFUS. Results: Our study showed that all seven patients tolerated the MRgFUSmore » treatment well. No skin toxicity or other complications were observed. The pain score (0–10) using the visual analog scale (VAS) was significantly reduced from 8.0 ± 1.1 before treatment to 4.7 ± 3.0, 3.0 ± 1.5, 3.2 ± 2.8 and 3.4 ± 1.5 at one day, one month, two months and three months after the MRgFUS treatment, respectively. Conclusion: We demonstrated that with the appropriate QA procedures, MRgFUS is a safe, effective and noninvasive treatment modality for palliation of bone metastases.« less

Authors:
; ; ; ;  [1]
  1. Fox Chase Cancer Center, Philadelphia, PA (United States)
Publication Date:
OSTI Identifier:
22334054
Resource Type:
Journal Article
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 41; Journal Issue: 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
Country of Publication:
United States
Language:
English
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; 60 APPLIED LIFE SCIENCES; ACCIDENTS; BURNS; COMPUTER CODES; COOLING SYSTEMS; HEAD; IMAGES; LIGHT BULBS; METASTASES; NMR IMAGING; PAIN; PATIENTS; QUALITY ASSURANCE; SAFETY; SKELETON; SKIN; TOXICITY; TRANSDUCERS

Citation Formats

Chen, L, Chen, X, Wang, B, Gupta, R, and Ma, C. SU-E-J-162: Quality Assurance Procedures for MR Guided Focused Ultrasound Treatment of Bone Metastasis. United States: N. p., 2014. Web. doi:10.1118/1.4888215.
Chen, L, Chen, X, Wang, B, Gupta, R, & Ma, C. SU-E-J-162: Quality Assurance Procedures for MR Guided Focused Ultrasound Treatment of Bone Metastasis. United States. https://doi.org/10.1118/1.4888215
Chen, L, Chen, X, Wang, B, Gupta, R, and Ma, C. 2014. "SU-E-J-162: Quality Assurance Procedures for MR Guided Focused Ultrasound Treatment of Bone Metastasis". United States. https://doi.org/10.1118/1.4888215.
@article{osti_22334054,
title = {SU-E-J-162: Quality Assurance Procedures for MR Guided Focused Ultrasound Treatment of Bone Metastasis},
author = {Chen, L and Chen, X and Wang, B and Gupta, R and Ma, C},
abstractNote = {Purpose: The purpose of this work is to develop and verify our quality assurance (QA) procedures to ensure the safety and efficacy of MR-guided focused ultrasound (MRgFUS) treatment of bone metastases. Methods: A practical QA program was developed. Monthly and daily QA (DQA) procedures were performed. The major QA items included the checks of the machine hardware, software and patient safety features. Briefly, these checks/tests include: 1) the cooling system reservoir and treatment table; 2) power to the treatment table; 3) the MR coil; 4) the transducer position with MRI; 5) image display on the treatment work station; 6) the effective focal spot in 3 directions using MR thermometry; and 7) all the safety devices including a sonication lamp, and the emergency stop-sonication switches. In order to avoid patient skin burn, it is important to remove gas bubbles in the interfaces between the treatment table and the gel pad, and the gel pad and patients skin during the patient setup. Our QA procedures have been verified and evaluated through patient treatments. Seven patients with scapula, humeral head, sacrum, ilium, pubic ramus and acetabular bone metastases were treated using MRgFUS. Results: Our study showed that all seven patients tolerated the MRgFUS treatment well. No skin toxicity or other complications were observed. The pain score (0–10) using the visual analog scale (VAS) was significantly reduced from 8.0 ± 1.1 before treatment to 4.7 ± 3.0, 3.0 ± 1.5, 3.2 ± 2.8 and 3.4 ± 1.5 at one day, one month, two months and three months after the MRgFUS treatment, respectively. Conclusion: We demonstrated that with the appropriate QA procedures, MRgFUS is a safe, effective and noninvasive treatment modality for palliation of bone metastases.},
doi = {10.1118/1.4888215},
url = {https://www.osti.gov/biblio/22334054}, journal = {Medical Physics},
issn = {0094-2405},
number = 6,
volume = 41,
place = {United States},
year = {Sun Jun 01 00:00:00 EDT 2014},
month = {Sun Jun 01 00:00:00 EDT 2014}
}