skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-D-19A-03: Monte Carlo Investigation of the Mobetron to Perform Modulated Electron Beam Therapy

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4887866· OSTI ID:22333972
 [1];  [2]; ;  [3];  [2]
  1. Ain Shams University, Cairo (Egypt)
  2. Fox Chase Cancer Center, Philadelphia, PA (United States)
  3. King Saud University Hospitals, Riyadh (Saudi Arabia)

Purpose: Modulated electron radiotherapy (MERT) has been proposed as a mean of delivering conformal dose to shallow tumors while sparing distal structures and surrounding tissues. In intraoperative radiotherapy (IORT) utilizing Mobetron, an applicator is placed as closely as possible to the suspected cancerous tissues to be treated. In this study we investigate the characteristics of Mobetron electron beams collimated by an in-house prospective electron multileaf collimator (eMLC) and its feasibility for MERT. Methods: IntraOp Mobetron™ dedicated to perform radiotherapy during surgery was used in the study. It provides several energies (6, 9 and 12 MeV). Dosimetry measurements were performed to obtain percentage depth dose curves (PDD) and profiles for a 10-cm diameter applicator using the PTW MP3/XS 3D-scanning system and the semiflex ion chamber. MCBEAM/MCSIM Monte Carlo codes were used for the treatment head simulation and phantom dose calculation. The design of an electron beam collimation by an eMLC attached to the Mobetron head was also investigated using Monte Carlo simulations. Isodose distributions resulting from eMLC collimated beams were compared to that collimated using cutouts. The design for our Mobetron eMLC is based on our previous experiences with eMLCs designed for clinical linear accelerators. For Mobetron the eMLC is attached to the end of a spacer-mounted rectangular applicator at 50 cm SSD. Steel will be used as the leaf material because other materials would be toxic and will not be suitable for intraoperative applications. Results: Good agreement (within 2%) was achieved between measured and calculated PDD curves and profiles for all available energies. Dose distributiosn provided by the eMLC showed reasonable agreement (∼3%/1mm) with those obtained by conventional cutouts. Conclusion: Monte Carlo simulations are capable of modeling Mobetron electron beams with a reliable accuracy. An eMLC attached to the Mobteron treatment head will allow better treatment options with those machines.

OSTI ID:
22333972
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English