Title: SU-E-J-24: An Evaluation of the Stability of Image Quality Parameters of Varian On-Board Imaging (OBI) and EPID Imaging Systems

Purpose: Quality assurance of the image quality for image guided localization systems is crucial to ensure accurate visualization and localization of target volumes. In this study, the long term stability of selected image parameters was assessed and evaluated for CBCT mode, planar radiographic kV mode and the radiographic MV EPID mode. Methods: The CATPHAN, QckV-1 and QC-3 phantoms were used to evaluate the image quality parameters. The planar radiographic images were analyzed in PIPSpro™ with spatial resolution (f30, f40, f50) being recorded. For OBI CBCT, High quality head Full-Fan acquisition and Pelvis Half-Fan acquisition modes were evaluated for Uniformity, Noise, Spatial Resolution, HU constancy and geometric distortion. Dose and kVp for the OBI were recorded using the Unfors RaySafe Xi system with the R/F High Detector for planar kV and the CT detector for CBCT. Dose for the MV EPID was recorded using a PTW975 Semiflex Ion Chamber, PTW Unidos electrometer and SolidWater™. Results: For each metric, values were normalized to the mean and the standard deviations were recorded. For the planar radiographic spatial resolution the f30, f40, f50 were 0.015, 0.008, 0.004 and 0.006, 0.009, 0.018 for the kV and MV, respectively. The standard deviation of the dose formore » kV was 0.010 and 0.005 for the MV. The standard deviations for Full and half fan were averaged together and the following standard deviations for each metric were recorded: 0.075(uniformity), 0.071(noise), 0.006(AP-Geometric Distortion), 0.005(LAT-Geometric Distortion), 0.058(mean slice thickness), 0.098(f30),0.101(f40),0.124(f50), 0.031(Lung/PMP-HU constancy), 0.063(Water/poly-HU constancy), 0.015(Bone/Derlin-HU constancy),0.006(Dose-Center), 0.004(Dose-Periphery). Using these, tolerances can be reported as a warning and action threshold of 1σ and 2σ. Conclusion: A study was performed to assess the stability of the basic image quality parameters recommended by TG-142 for the Varian OBI and EPID Imaging systems. The two systems show consistent imaging and dosimetric properties over the evaluated time frame.« less
Authors:
; ;  [1]
  1. University of Texas HSC SA, San Antonio, TX (United States)
Publication Date:
OSTI Identifier:
22325198
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 41; Journal Issue: 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; 60 APPLIED LIFE SCIENCES; BLOWERS; ELECTROMETERS; FLUORINE 30; IMAGES; IONIZATION CHAMBERS; LUNGS; METRICS; NOISE; PELVIS; PHANTOMS; QUALITY ASSURANCE; RADIATION DOSES; SPATIAL RESOLUTION; STABILITY; STANDARDS; TOLERANCE