SU-E-I-17: Comparison of Two Novel Algorithms for the Modulation Transfer Function of CT Using a Simple Cylindrical Phantom
- Pusan National University, Busan (Korea, Republic of)
- Pusan National University Yangsan Hospital, Yangsan (Korea, Republic of)
Purpose: To compare and analyze two novel algorithms for the assessment of modulation transfer functions (MTFs) of computed tomography (CT) systems using a simple acrylic cylindrical phantom Method and Materials: Images of the acrylic cylindrical phantom were acquired by a GE LightSpeed 16 RT (GE Healthcare, Milwaukee, WI) using 120 kVp, 330 mA, 2.5 mm slice thickness, 10 cm field-of view (FOV), four reconstruction kernels (e.g. standard, soft, detail, bone, and lung). Two different algorithms were used to analyze images for MTF assessment. First, Richard et al. suggested a task-based MTF assessment method through an edge spread function (ESF) which described pixel intensities as a function of distance from the center. The MTF was obtained as the absolute value of Fourier transform of the differentiated ESF. Second, Ohkubo et al. devised an effective method to determine the point spread function (PSF) of CT system accompanied with verification. The line spread function (LSF), which was the one-dimensional integration of the PSF, was used to obtain the MTF. We validated the reliability of two above-mentioned methods through the comparison with a conventional method using a thin tungsten wire phantom. Results: The measured MTFs by two methods were mostly similar each other for standard, soft, and detail kernels. In 0.6 lp/mm, the MTF difference between two methods were 0.012(standard), 0.004(soft), and 0.037(detail). They also coincided with the MTF by the conventional method well. However, there were considerable distinctions for bone and lung kernels containing edge enhancement that might cause undershoots near the peak of the LSF. Conclusions: We compared two novel methods to assess task-based MTFs for clinical CT systems especially using a simple acrylic cylindrical phantom with high-convenience and low-cost, and validated them against a conventional method. This work can provide a practical solution to users for the quality assurance of CT.
- OSTI ID:
- 22325136
- Journal Information:
- Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-G-206-02: Impact of Focal Spot Sizes On CT Image Quality
Model-based PSF and MTF estimation and validation from skeletal clinical CT images
Related Subjects
60 APPLIED LIFE SCIENCES
ALGORITHMS
CHARGES
COMPARATIVE EVALUATIONS
COMPUTERIZED TOMOGRAPHY
CYLINDRICAL CONFIGURATION
FOURIER TRANSFORMATION
KERNELS
LUNGS
MATHEMATICAL SOLUTIONS
MODULATION
ONE-DIMENSIONAL CALCULATIONS
PHANTOMS
QUALITY ASSURANCE
SKELETON
SOLUTIONS
THICKNESS
TRANSFER FUNCTIONS
TUNGSTEN
VERIFICATION