Title: Evaluating the purity of a {sup 57}Co flood source by PET

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4897611· OSTI ID:22320358
 [1]
  1. Department of Nuclear Medicine, Cleveland Clinic, Cleveland, Ohio 44195 (United States)

Purpose: Flood sources of {sup 57}Co are commonly used for quality control of gamma cameras. Flood uniformity may be affected by the contaminants {sup 56}Co and {sup 58}Co, which emit higher energy photons. Although vendors specify a maximum combined {sup 56}Co and {sup 58}Co activity, a convenient test for flood source purity that is feasible in a clinical environment would be desirable. Methods: Both {sup 56}Co and {sup 58}Co emit positrons with branching 19.6% and 14.9%, respectively. As is known from {sup 90}Y imaging, a positron emission tomography (PET) scanner is capable of quantitatively imaging very weak positron emission in a high single-photon background. To evaluate this approach, two {sup 57}Co flood sources were scanned with a clinical PET/CT multiple times over a period of months. The {sup 56}Co and {sup 58}Co activity was clearly visible in the reconstructed PET images. Total impurity activity was quantified from the PET images after background subtraction of prompt gamma coincidences. Results: Time-of-flight PET reconstruction was highly beneficial for accurate image quantification. Repeated measurements of the positron-emitting impurities showed excellent agreement with an exponential decay model. For both flood sources studied, the fit parameters indicated a zero intercept and a decay half-life consistent with a mixture of {sup 56}Co and {sup 58}Co. The total impurity activity at the reference date was estimated to be 0.06% and 0.07% for the two sources, which was consistent with the vendor’s specification of <0.12%. Conclusions: The robustness of the repeated measurements and a thorough analysis of the detector corrections and physics suggest that the accuracy is acceptable and that the technique is feasible. Further work is needed to validate the accuracy of this technique with a calibrated high resolution gamma spectrometer as a gold standard, which was not available for this study, and for other PET detector models.

OSTI ID:
22320358
Journal Information:
Medical Physics, Vol. 41, Issue 11; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English