Title: Note: A novel normalization scheme for laser-based plasma x-ray sources

A kHz repetition rate laser pump-X-ray probe system for ultrafast X-ray diffraction is set up based on a laser-driven plasma X-ray source. A simple and reliable normalization approach has been developed to minimize the impact of large X-ray pulse intensity fluctuation on data quality. It utilizes one single X-ray area detector to record both sample and reference signals simultaneously. Performance of this novel normalization method is demonstrated in reflectivity oscillation measurement of a superlattice sample at sub-ps resolution.
Authors:
; ;  [1] ;  [2]
  1. Beijing Synchrotron Radiation Facility, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049 (China)
  2. Institute of Physics, Chinese Academy of Sciences, Beijing 100090 (China)
Publication Date:
OSTI Identifier:
22314289
Resource Type:
Journal Article
Resource Relation:
Journal Name: Review of Scientific Instruments; Journal Volume: 85; Journal Issue: 9; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; FLUCTUATIONS; KHZ RANGE; LASERS; OSCILLATIONS; PLASMA; PROBES; PULSES; PUMPING; REFLECTIVITY; RESOLUTION; X-RAY DIFFRACTION; X-RAY SOURCES