Title: Coded aperture imaging for fluorescent x-rays

We employ a coded aperture pattern in front of a pixilated charge couple device detector to image fluorescent x-rays (6–25 KeV) from samples irradiated with synchrotron radiation. Coded apertures encode the angular direction of x-rays, and given a known source plane, allow for a large numerical aperture x-ray imaging system. The algorithm to develop and fabricate the free standing No-Two-Holes-Touching aperture pattern was developed. The algorithms to reconstruct the x-ray image from the recorded encoded pattern were developed by means of a ray tracing technique and confirmed by experiments on standard samples.
Authors:
 [1] ;  [2] ; ; ;  [1]
  1. Lawrence Berkeley National Laboratories, Berkeley, California 94720 (United States)
  2. (United States)
Publication Date:
OSTI Identifier:
22308811
Resource Type:
Journal Article
Resource Relation:
Journal Name: Review of Scientific Instruments; Journal Volume: 85; Journal Issue: 6; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; ALGORITHMS; APERTURES; IMAGES; KEV RANGE; SYNCHROTRON RADIATION; X-RAY FLUORESCENCE ANALYSIS