Title: Polarimetry diagnostic on OMEGA EP using a 10-ps, 263-nm probe beam

A polarimetry diagnostic was built and characterized for magnetic-field measurements in laser-plasma experiments on the OMEGA EP laser. This diagnostic was built into the existing 4ω (263-nm) probe system that employs a 10-ps laser pulse collected with an f/4 imaging system. The diagnostic measures the rotation of the probe beam's polarization. The polarimeter uses a Wollaston prism to split the probe beam into orthogonal polarization components. Spatially localized intensity variations between images indicate polarization rotation. Magnetic fields can be calculated by combining the polarimetry data with the measured plasma density profile obtained from angular filter refractometry.
Authors:
; ; ; ; ;  [1]
  1. Laboratory for Laser Energetics, University of Rochester, Rochester, New York 14623 (United States)
Publication Date:
OSTI Identifier:
22308650
Resource Type:
Journal Article
Resource Relation:
Journal Name: Review of Scientific Instruments; Journal Volume: 85; Journal Issue: 11; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; BEAMS; LASER-PRODUCED PLASMA; MAGNETIC FIELDS; PLASMA DENSITY; PLASMA RADIAL PROFILES; POLARIMETERS; POLARIMETRY; POLARIZATION; PRISMS; PULSES; REFRACTION; ROTATION