skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: The development of in-situ calibration method for divertor IR thermography in ITER

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4894044· OSTI ID:22308283

For the development of the calibration method of the emissivity in IR light on the divertor plate in ITER divertor IR thermography system, the laboratory experiments have been performed by using IR instruments. The calibration of the IR camera was performed by the plane black body in the temperature of 100–600 degC. The radiances of the tungsten heated by 280 degC were measured by the IR camera without filter (2.5–5.1 μm) and with filter (2.95 μm, 4.67 μm). The preliminary data of the scattered light of the laser of 3.34 μm that injected into the tungsten were acquired.

OSTI ID:
22308283
Journal Information:
AIP Conference Proceedings, Vol. 1612, Issue 1; Conference: International conference on fusion reactor diagnostics, Varenna (Italy), 9-13 Sep 2013; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
Country of Publication:
United States
Language:
English

Similar Records

Divertor IR thermography on Alcator C-Mod
Journal Article · Fri Oct 15 00:00:00 EDT 2010 · Review of Scientific Instruments · OSTI ID:22308283

Development of a mirror-based endoscope for divertor spectroscopy on JET with the new ITER-like wall (invited)
Journal Article · Mon Oct 15 00:00:00 EDT 2012 · Review of Scientific Instruments · OSTI ID:22308283

Gamma-irradiation tests of IR optical fibres for ITER thermography--a case study
Journal Article · Wed Mar 12 00:00:00 EDT 2008 · AIP Conference Proceedings · OSTI ID:22308283