Title: The development of in-situ calibration method for divertor IR thermography in ITER

For the development of the calibration method of the emissivity in IR light on the divertor plate in ITER divertor IR thermography system, the laboratory experiments have been performed by using IR instruments. The calibration of the IR camera was performed by the plane black body in the temperature of 100–600 degC. The radiances of the tungsten heated by 280 degC were measured by the IR camera without filter (2.5–5.1 μm) and with filter (2.95 μm, 4.67 μm). The preliminary data of the scattered light of the laser of 3.34 μm that injected into the tungsten were acquired.
Authors:
; ; ; ;  [1]
  1. Japan Atomic Energy Agency (Japan)
Publication Date:
OSTI Identifier:
22308283
Resource Type:
Journal Article
Resource Relation:
Journal Name: AIP Conference Proceedings; Journal Volume: 1612; Journal Issue: 1; Conference: International conference on fusion reactor diagnostics, Varenna (Italy), 9-13 Sep 2013; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; 71 CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS; CALIBRATION; CAMERAS; DIVERTORS; EMISSIVITY; FILTERS; ITER TOKAMAK; LASERS; THERMOGRAPHY; TUNGSTEN