skip to main content

Title: Precise method of compensating radiation-induced errors in a hot-cathode-ionization gauge with correcting electrode

To compensate pressure-measurement errors caused by a synchrotron radiation environment, a precise method using a hot-cathode-ionization-gauge head with correcting electrode, was developed and tested in a simulation experiment with excess electrons in the SPring-8 storage ring. This precise method to improve the measurement accuracy, can correctly reduce the pressure-measurement errors caused by electrons originating from the external environment, and originating from the primary gauge filament influenced by spatial conditions of the installed vacuum-gauge head. As the result of the simulation experiment to confirm the performance reducing the errors caused by the external environment, the pressure-measurement error using this method was approximately less than several percent in the pressure range from 10{sup −5} Pa to 10{sup −8} Pa. After the experiment, to confirm the performance reducing the error caused by spatial conditions, an additional experiment was carried out using a sleeve and showed that the improved function was available.
Authors:
;  [1]
  1. Japan Synchrotron Radiation Research Institute, SPring-8, Kohto 1-1-1, Sayo, Hyogo 679-5198 (Japan)
Publication Date:
OSTI Identifier:
22307963
Resource Type:
Journal Article
Resource Relation:
Journal Name: AIP Conference Proceedings; Journal Volume: 1618; Journal Issue: 1; Conference: ICCMSE 2014: International conference on computational methods in science and engineering 2014, Athens (Greece), 4-7 Apr 2014; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
71 CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS; ACCURACY; APPROXIMATIONS; CATHODES; COMPUTERIZED SIMULATION; ELECTRONS; ERRORS; IONIZATION GAGES; MEASURING METHODS; PERFORMANCE; PRESSURE MEASUREMENT; SYNCHROTRON RADIATION