Precise method of compensating radiation-induced errors in a hot-cathode-ionization gauge with correcting electrode
- Japan Synchrotron Radiation Research Institute, SPring-8, Kohto 1-1-1, Sayo, Hyogo 679-5198 (Japan)
To compensate pressure-measurement errors caused by a synchrotron radiation environment, a precise method using a hot-cathode-ionization-gauge head with correcting electrode, was developed and tested in a simulation experiment with excess electrons in the SPring-8 storage ring. This precise method to improve the measurement accuracy, can correctly reduce the pressure-measurement errors caused by electrons originating from the external environment, and originating from the primary gauge filament influenced by spatial conditions of the installed vacuum-gauge head. As the result of the simulation experiment to confirm the performance reducing the errors caused by the external environment, the pressure-measurement error using this method was approximately less than several percent in the pressure range from 10{sup −5} Pa to 10{sup −8} Pa. After the experiment, to confirm the performance reducing the error caused by spatial conditions, an additional experiment was carried out using a sleeve and showed that the improved function was available.
- OSTI ID:
- 22307963
- Journal Information:
- AIP Conference Proceedings, Vol. 1618, Issue 1; Conference: ICCMSE 2014: International conference on computational methods in science and engineering 2014, Athens (Greece), 4-7 Apr 2014; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
- Country of Publication:
- United States
- Language:
- English
Similar Records
Laser - Polarized HE-3 Target Used for a Precision Measurement of the Neutron Spin Structure
Scheme for precise correction of orbit variation caused by dipole error field of insertion device