Development of a dual MCP framing camera for high energy x-rays
- Lawrence Livermore National Laboratory, Livermore, California 94550 (United States)
Recently developed diagnostic techniques at LLNL require recording backlit images of extremely dense imploded plasmas using hard x-rays, and demand the detector to be sensitive to photons with energies higher than 50 keV [R. Tommasini et al., Phys. Phys. Plasmas 18, 056309 (2011); G. N. Hall et al., “AXIS: An instrument for imaging Compton radiographs using ARC on the NIF,” Rev. Sci. Instrum. (these proceedings)]. To increase the sensitivity in the high energy region, we propose to use a combination of two MCPs. The first MCP is operated in a low gain regime and works as a thick photocathode, and the second MCP works as a high gain electron multiplier. We tested the concept of this dual MCP configuration and succeeded in obtaining a detective quantum efficiency of 4.5% for 59 keV x-rays, 3 times larger than with a single plate of the thickness typically used in NIF framing cameras.
- OSTI ID:
- 22303690
- Journal Information:
- Review of Scientific Instruments, Vol. 85, Issue 11; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Sensitivity of NIF-scale backlit thin shell implosions to hohlraum symmetry in the foot of the ignition drive pulse
Polar direct drive: Proof-of-principle experiments on OMEGA and prospects for ignition on the National Ignition Facility