Title: Validation experiment of a numerically processed millimeter-wave interferometer in a laboratory

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4893430· OSTI ID:22303686
;  [1];  [2]; ;  [3]; ;  [4];  [5]
  1. Department of Information Electronics, Fukuoka Institute of Technology, Fukuoka 811-0295 (Japan)
  2. Art, Science and Technology Center for Cooperative Research, Kyushu University, Kasuga, Fukuoka 816-0811 (Japan)
  3. Plasma Research Center, University of Tsukuba, Tsukuba, Ibaraki 305-8577 (Japan)
  4. National Institute for Fusion Science, Toki, Gifu 509-5202 (Japan)
  5. Tokyo University of Agriculture and Technology, Koganei, Tokyo 184-8588 (Japan)

We propose a new interferometer system for density profile measurements. This system produces multiple measurement chords by a leaky-wave antenna driven by multiple frequency inputs. The proposed system was validated in laboratory evaluation experiments. We confirmed that the interferometer generates a clear image of a Teflon plate as well as the phase shift corresponding to the plate thickness. In another experiment, we confirmed that quasi-optical mirrors can produce multiple measurement chords; however, the finite spot size of the probe beam degrades the sharpness of the resulting image.

OSTI ID:
22303686
Journal Information:
Review of Scientific Instruments, Vol. 85, Issue 11; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English

Similar Records

Millimeter-wave phase-imaging interferometer for the GAMMA 10 tandem mirror
Journal Article · Sun Dec 01 00:00:00 EST 1991 · Review of Scientific Instruments; (United States) · OSTI ID:22303686

Design and implementation of a 150 GHz single-channel millimeter wave interferometer on Joint TEXT tokamak
Journal Article · Mon Apr 15 00:00:00 EDT 2013 · Review of Scientific Instruments · OSTI ID:22303686

Multichannel microwave interferometer with an antenna switching system for electron density measurement in a laboratory plasma experiment
Journal Article · Sat Feb 15 00:00:00 EST 2014 · Review of Scientific Instruments · OSTI ID:22303686