skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: A high-sensitivity push-pull magnetometer

Journal Article · · Applied Physics Letters
DOI:https://doi.org/10.1063/1.4861458· OSTI ID:22275821
; ; ;  [1]
  1. Department of Physics, University of Fribourg, Fribourg 1700 (Switzerland)

We describe our approach to atomic magnetometry based on the push-pull optical pumping technique. Cesium vapor is pumped and probed by a resonant laser beam whose circular polarization is modulated synchronously with the spin evolution dynamics induced by a static magnetic field. The magnetometer is operated in a phase-locked loop, and it has an intrinsic sensitivity below 20fT/√(Hz), using a room temperature paraffin-coated cell. We use the magnetometer to monitor magnetic field fluctuations with a sensitivity of 300fT/√(Hz)

OSTI ID:
22275821
Journal Information:
Applied Physics Letters, Vol. 104, Issue 2; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0003-6951
Country of Publication:
United States
Language:
English

Similar Records

In situ magnetic compensation for potassium spin-exchange relaxation-free magnetometer considering probe beam pumping effect
Journal Article · Sun Jun 15 00:00:00 EDT 2014 · Review of Scientific Instruments · OSTI ID:22275821

Dual channel self-oscillating optical magnetometer
Journal Article · Fri May 15 00:00:00 EDT 2009 · Journal of the Optical Society of America. Part B, Optical Physics · OSTI ID:22275821

Sensitive YBa sub 2 Cu sub 3 O sub 7 minus x thin-film magnetometer
Journal Article · Mon Aug 19 00:00:00 EDT 1991 · Applied Physics Letters; (United States) · OSTI ID:22275821