Title: Performance study of polycrystalline CVD diamond detectors for fast neutron monitoring

Diamond detectors using polycrystalline CVD diamond substrates of thickness 300μm and 100μm were fabricated for fast neutron monitoring application.. The characterization of detectors was carried out using various tests such as leakage current, capacitance and alpha particle response. The performance of detectors was evaluated for fast neutrons at different neutron yields. The results presented in this work demonstrate that the diamond detectors will be suitable for monitoring fast neutrons.
Authors:
; ;  [1]
  1. Electronics Division, Bhabha Atomic Research Centre, Mumbai-400085 (India)
Publication Date:
OSTI Identifier:
22269524
Resource Type:
Journal Article
Resource Relation:
Journal Name: AIP Conference Proceedings; Journal Volume: 1591; Journal Issue: 1; Conference: 58. DAE solid state physics symposium 2013, Patiala, Punjab (India), 17-21 Dec 2013; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; 75 CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY; CAPACITANCE; CHEMICAL VAPOR DEPOSITION; DIAMONDS; FAST NEUTRONS; LEAKAGE CURRENT; NEUTRON DETECTION; NEUTRON DETECTORS; PERFORMANCE; PERFORMANCE TESTING; POLYCRYSTALS; RADIATION MONITORING; SUBSTRATES; THICKNESS