skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: X-ray source considerations in operation of digital detector arrays

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4865045· OSTI ID:22263774
;  [1]
  1. Iowa State University, Center for NDE, 1915 Scholl Road, Ames, IA 50011 (United States)

Digital Detector Arrays (DDA) are increasingly replacing film in radiography applications. Standards exist for characterizing the performance of these detectors, and for using them in specific inspections. We have observed that the selection of the x-ray source to use with these detectors can also have a significant influence on the performance. We look at differences between standard, and micro-focus x-ray tubes, and end-window vs. side-window micro-focus tubes. We find that for best results, one must calibrate the DDA for the source settings used during an inspection. This is particularly true for variable-focus sources.

OSTI ID:
22263774
Journal Information:
AIP Conference Proceedings, Vol. 1581, Issue 1; Conference: 40. annual review of progress in quantitative nondestructive evaluation, Baltimore, MD (United States), 21-26 Jul 2013, 10. international conference on Barkhausen noise and micromagnetic testing, Baltimore, MD (United States), 21-26 Jul 2013; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
Country of Publication:
United States
Language:
English