skip to main content

Title: Establishing an optimized patient-specific verification program for volumetric modulated arc therapy

Quality assurance (QA) of volumetric modulated arc therapy (VMAT) increases the workload significantly. We compared the results from 4 verification methods to establish an efficient VMAT QA. Planning for VMAT treatments was carried out for 40 consecutive patients. Pretreatment verifications were carried out with ion chamber array Physikalish-Technische Werkstätten (PTW729), electronic portal dosimetry (EPID), ion chamber measurements, and independent dose calculation with Diamond program. 2D analyses were made using the gamma analysis (3 mm distance to agreement and 3% dose difference relative to maximum, 10% dose threshold). Average point dose difference calculated by Eclipse relative to ion chamber measurements and Diamond were 0.1%±0.9% and 0.6%±2.2%, respectively. Average pass rate for PTW729 was 99.2%±1.9% and 98.3%±1.3% for EPID. The total required time (linac occupancy time given in parentheses) for each QA method was: PTW729 43.5 minutes (26.5 minutes), EPID 14.5 minutes (2.5 minutes), ion chamber 34.5 minutes (26.5 minutes), and Diamond 12.0 minutes (0 minute). The results were consistent and allowed us to establish an optimized protocol, considering safety and accuracy as well as workload, consisting of 2 verification methods: EPID 2D analysis and independent dose calculation.
Authors:
 [1] ; ;  [1]
  1. Department of Medical Physics, Hospital Universitario Santa Lucía, Cartagena (Spain)
Publication Date:
OSTI Identifier:
22262823
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Dosimetry; Journal Volume: 38; Journal Issue: 3; Other Information: Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
61 RADIATION PROTECTION AND DOSIMETRY; ACCURACY; DOSIMETRY; EXTERNAL BEAM RADIATION THERAPY; IONIZATION CHAMBERS; LINEAR ACCELERATORS; PLANNING; QUALITY ASSURANCE; RADIATION DOSES; VERIFICATION