Title: OSLD energy response performance and dose accuracy at 24 - 1250 keV: Comparison with TLD-100H and TLD-100

OSLD was evaluated in terms of energy response and accuracy of the measured dose in comparison with TLD-100H and TLD-100. The OSLD showed a better energy response performance for H{sub p}(10) whereas for H{sub p}(0.07), TLD-100H is superior than the others. The OSLD dose accuracy is comparable with the other two dosimeters since it fulfilled the requirement of the ICRP trumpet graph analysis.