Title: The motional Stark effect polarimeter in the HL-2A tokamak

A 7-channel motional Stark effect polarimeter based on four polarizers and a spectrometer has been developed in the HL-2A tokamak, which is the first time successful utilizing this kind of polarimeter on a tokamak. The accuracy of the angle can reach ±0.25° in the calibration experiments. Pilot experiments of measuring the magnetic pitch angle have been successfully carried out in the weak motional Stark effect plasma discharge with toroidal magnetic field of ∼1.3 T and beam energy of ∼25 keV/amu. The pitch angles of magnetic field are obtained for 7 spatial points covering 24 cm along major radius with time resolution of 40 ms; the profiles of safety factor are obtained by combining with the Equilibrium and Reconstruction Fitting Code. The core value of safety factor (q) is less than 1 during the sawtooth oscillation and the position of q = 1 surface is well consistent with the results measured by soft X-ray array.
Authors:
; ; ; ; ; ; ; ; ; ; ;  [1]
  1. Southwestern Institute of Physics, Chengdu, Sichuan 610041 (China)
Publication Date:
OSTI Identifier:
22254877
Resource Type:
Journal Article
Resource Relation:
Journal Name: Review of Scientific Instruments; Journal Volume: 85; Journal Issue: 5; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
70 PLASMA PHYSICS AND FUSION TECHNOLOGY; ACCURACY; CALIBRATION; HL-2A TOKAMAK; MAGNETIC FIELDS; PLASMA; POLARIMETERS; SAWTOOTH OSCILLATIONS; SOFT X RADIATION; SPECTROMETERS; STARK EFFECT; TIME RESOLUTION