skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Calibration of a 32 channel electron cyclotron emission radiometer on the HL-2A tokamak

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4866640· OSTI ID:22253498
; ; ; ; ; ; ; ;  [1]
  1. Southwestern Institute of Physics, P. O. Box 432, Chengdu 610041 (China)

A novel 32-channel electron cyclotron emission radiometer has been designed and tested for the measurement of electron temperature profiles on the HL-2A tokamak. This system is based on the intermediate frequency filter detection technique, and has the features of wide working frequency range and high spatial resolution. Two relative calibration methods have been investigated: sweeping the toroidal magnetic field and hopping the output frequency of the local oscillator. Preliminary results show that both methods can ensure reasonable profiles.

OSTI ID:
22253498
Journal Information:
Review of Scientific Instruments, Vol. 85, Issue 2; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English