Title: Note: Portable rare-earth element analyzer using pyroelectric crystal

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4846635· OSTI ID:22251220
; ; ;  [1]
  1. Department of Materials Science and Engineering, Kyoto University, Sakyo, Kyoto 606-8501 (Japan)

We report a portable rare-earth element analyzer with a palm-top size chamber including the electron source of a pyroelectric crystal and the sample stage utilizing cathodoluminescence (CL) phenomenon. The portable rare-earth element analyzer utilizing CL phenomenon is the smallest reported so far. The portable rare-earth element analyzer detected the rare-earth elements Dy, Tb, Er, and Sm of ppm order in zircon, which were not detected by scanning electron microscopy-energy dispersive X-ray spectroscopy analysis. We also performed an elemental mapping of rare-earth elements by capturing a CL image using CCD camera.

OSTI ID:
22251220
Journal Information:
Review of Scientific Instruments, Vol. 84, Issue 12; Other Information: (c) 2013 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English