Title: A dosimetric characterization of a novel linear accelerator collimator

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4866228· OSTI ID:22250943
;  [1];  [2]
  1. Leeds Cancer Centre, Bexley Wing, St. James’ University Hospital, Leeds LS9 7TF (United Kingdom)
  2. Institute of Medical Physics, School of Physics, University of Sydney, Sydney NSW 2006, Australia and Division of Medical Physics, Leeds Institute of Genetics, Health and Therapeutics, University of Leeds, Leeds LS2 9JT (United Kingdom)

Purpose: The aim of this work is to characterize a new linear accelerator collimator which contains a single pair of sculpted diaphragms mounted orthogonally to a 160 leaf multileaf collimator (MLC). The diaphragms have “thick” regions providing full attenuation and “thin” regions where attenuation is provided by both the leaves and the diaphragm. The leaves are mounted on a dynamic leaf guide allowing rapid leaf motion and leaf travel over 350 mm. Methods: Dosimetric characterization, including assessment of leaf transmission, leaf tip transmission, penumbral width, was performed in a plotting tank. Head scatter factor was measured using a mini-phantom and the effect of leaf guide position on output was assessed using a water phantom. The tongue and groove effect was assessed using multiple exposures on radiochromic film. Leaf reproducibility was assessed from portal images of multiple abutting fields. Results: The maximum transmission through the multileaf collimator is 0.44% at 6 MV and 0.52% at 10 MV. This reduced to 0.22% and 0.27%, respectively, when the beam passes through the dynamic leaf guide in addition to the MLC. The maximum transmission through the thick part of the diaphragm is 0.32% and 0.36% at 6 and 10 MV. The combination of leaf and diaphragm transmission ranges from 0.08% to 0.010% at 6 MV and 0.10% to 0.14% depending on whether the shielding is through the thick or thin part of the diaphragm. The off-axis intertip transmission for a zero leaf gap is 2.2% at 6 and 10 MV. The leaf tip penumbra for a 100 × 100 mm field ranges from 5.4 to 4.3 mm at 6 and 10 MV across the full range of leaf motion when measured in the AB direction, which reduces to 4.0–3.4 mm at 6 MV and 4.5–3.8 mm at 10 MV when measured in the GT direction. For a 50 × 50 mm field, the diaphragm penumbra ranges from 4.3 to 3.7 mm at 6 MV and 4.5 to 4.1 mm at 10 MV in the AB direction and 3.7 to 3.2 mm at 6 MV and 4.2 to 3.7 mm when measured in the GT direction. The tongue and groove effect observed from exposure of a radiochromic film to two abutting fields is an underdose of 25%. The head scatter factor at both 6 and 10 MV is similar to that from the MLCi2 collimator to within 0.8%. The uncertainty in the leaf position reproducibility is 0.05 mm (2σ). Conclusions: The Agility collimator is a low leakage, high definition collimator where both the MLC and the sculpted diaphragm have been optimized for dynamic treatments.

OSTI ID:
22250943
Journal Information:
Medical Physics, Vol. 41, Issue 3; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

Dosimetric characterization of a multileaf collimator for a new four-dimensional image-guided radiotherapy system with a gimbaled x-ray head, MHI-TM2000
Journal Article · Wed Sep 15 00:00:00 EDT 2010 · Medical Physics · OSTI ID:22250943

Monte Carlo implementation, validation, and characterization of a 120 leaf MLC
Journal Article · Sat Oct 15 00:00:00 EDT 2011 · Medical Physics · OSTI ID:22250943

Characterization of a multileaf collimator system
Journal Article · Fri Jan 15 00:00:00 EST 1993 · International Journal of Radiation Oncology, Biology and Physics; (United States) · OSTI ID:22250943