skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Compensator models for fluence field modulated computed tomography

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4829513· OSTI ID:22230761
 [1];  [1]
  1. Department of Medical Biophysics, University of Toronto, Toronto, Ontario M5G 2M9 (Canada)

Purpose: Fluence field modulated computed tomography (FFMCT) presents a novel approach for acquiring CT images, whereby a patient model guides dynamically changing fluence patterns in an attempt to achieve task-based, user-prescribed, regional variations in image quality, while also controlling dose to the patient. This work aims to compare the relative effectiveness of FFMCT applied to different thoracic imaging tasks (routine diagnostic CT, lung cancer screening, and cardiac CT) when the modulator is subject to limiting constraints, such as might be present in realistic implementations.Methods: An image quality plan was defined for a simulated anthropomorphic chest slice, including regions of high and low image quality, for each of the thoracic imaging tasks. Modulated fluence patterns were generated using a simulated annealing optimization script, which attempts to achieve the image quality plan under a global dosimetric constraint. Optimization was repeated under different types of modulation constraints (e.g., fixed or gantry angle dependent patterns, continuous or comprised of discrete apertures) with the most limiting case being a fixed conventional bowtie filter. For each thoracic imaging task, an image quality map (IQM{sub sd}) representing the regionally varying standard deviation is predicted for each modulation method and compared to the prescribed image quality plan as well as against results from uniform fluence fields. Relative integral dose measures were also compared.Results: Each IQM{sub sd} resulting from FFMCT showed improved agreement with planned objectives compared to those from uniform fluence fields for all cases. Dynamically changing modulation patterns yielded better uniformity, improved image quality, and lower dose compared to fixed filter patterns with optimized tube current. For the latter fixed filter cases, the optimal choice of tube current modulation was found to depend heavily on the task. Average integral dose reduction compared to a uniform fluence field ranged from 10% using a bowtie filter to 40% or greater using an idealized modulator.Conclusions: The results support that FFMCT may achieve regionally varying image quality distributions in good agreement with user-prescribed values, while limiting dose. The imposition of constraints inhibits dose reduction capacity and agreement with image quality plans but still yields significant improvement over what is afforded by conventional dose minimization techniques. These results suggest that FFMCT can be implemented effectively even when the modulator has limited modulation capabilities.

OSTI ID:
22230761
Journal Information:
Medical Physics, Vol. 40, Issue 12; Other Information: (c) 2013 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English