skip to main content

Title: Calculations of two new dose metrics proposed by AAPM Task Group 111 using the measurements with standard CT dosimetry phantoms

Purpose: AAPM Task Group 111 proposed to measure the equilibrium dose-pitch product D-caret{sub eq} for scan modes involving table translation and the midpoint dose D{sub L}(0) for stationary-table modes on the central and peripheral axes of sufficiently long (e.g., at least 40 cm) phantoms. This paper presents an alternative approach to calculate both metrics using the measurements of scanning the standard computed tomographic (CT) dosimetry phantoms on CT scanners.Methods: D-caret{sub eq} was calculated from CTDI{sub 100} and ε(CTDI{sub 100}) (CTDI{sub 100} efficiency), and D{sub L}(0) was calculated from D-caret{sub eq} and the approach to equilibrium function H(L) =D{sub L}(0)/D{sub eq}, where D{sub eq} was the equilibrium dose. CTDI{sub 100} may be directly obtained from several sources (such as medical physicist's CT scanner performance evaluation or the IMPACT CT patient dosimetry calculator), or be derived from CTDI{sub Vol} using the central to peripheral CTDI{sub 100} ratio (R{sub 100}). The authors have provided the required ε(CTDI{sub 100}) and H(L) data in two previous papers [X. Li, D. Zhang, and B. Liu, Med. Phys. 39, 901–905 (2012); and ibid. 40, 031903 (10pp.) (2013)]. R{sub 100} was assessed for a series of GE, Siemens, Philips, and Toshiba CT scanners with multiple settings of scanmore » field of view, tube voltage, and bowtie filter.Results: The calculated D{sub L}(0) and D{sub L}(0)/D{sub eq} in PMMA and water cylinders were consistent with the measurements on two GE CT scanners (LightSpeed 16 and VCT) by Dixon and Ballard [Med. Phys. 34, 3399–3413 (2007)], the measurements on a Siemens CT scanner (SOMATOM Spirit Power) by Descamps et al. [J. Appl. Clin. Med. Phys. 13, 293–302 (2012)], and the Monte Carlo simulations by Boone [Med. Phys. 36, 4547–4554 (2009)].Conclusions: D-caret{sub eq} and D{sub L}(0) can be calculated using the alternative approach. The authors have provided the required ε(CTDI{sub 100}) and H(L) data in two previous papers. R{sub 100} is presented for a majority of multidetector CT scanners currently on the market, and can be easily assessed for other CT scanners or operating conditions not covered in this study. The central to peripheral D{sub eq} ratio is about 1.50 and 1.12 times of R{sub 100} for the 32- and 16-cm diameter PMMA phantom, respectively.« less
Authors:
; ;  [1]
  1. Division of Diagnostic Imaging Physics and Webster Center for Advanced Research and Education in Radiation, Department of Radiology, Massachusetts General Hospital, Boston, Massachusetts 02114 (United States)
Publication Date:
OSTI Identifier:
22220521
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 40; Journal Issue: 8; Other Information: (c) 2013 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; 61 RADIATION PROTECTION AND DOSIMETRY; COMPUTERIZED SIMULATION; COMPUTERIZED TOMOGRAPHY; DOSIMETRY; MONTE CARLO METHOD; PATIENTS; PHANTOMS; PMMA; RADIATION DOSES