skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Realization of the WC-C peritectic fixed point at NIM and NMIJ

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4821409· OSTI ID:22218046
; ; ; ;  [1]; ; ;  [2]
  1. National Institute of Metrology (NIM), Beijing (China)
  2. National Metrology Institute of Japan, AIST (NMIJ, AIST), Tsukuba (Japan)

Three WC-C peritectic fixed point cells, constructed from different sources of tungsten with different nominal purities, were measured at NIM and NMIJ. The three cells were constructed at NMIJ by NIM and NMIJ staffs, and T{sub 90} values of the three cells were measured at NMIJ during the period 31 Aug. to 25 Dec. 2009. Thereafter, the three cells were then transported to NIM, and T{sub 90} values of these cells were measured from 7 Dec. 2011 to 9 Jan. 2012. The results showed that T{sub 90} values of the three cells measured at the two institutes agreed within 0.4 °C with the combined scale comparison uncertainty of 1.7 °C (k= 2). The main component of the uncertainty is not the uncertainty due to impurities of the cells but the scale uncertainty and the stability of the measurement system. From these results it can be concluded that the WC-C cell is stable enough to provide new means of international high-temperature scale comparison above 3000 K.

OSTI ID:
22218046
Journal Information:
AIP Conference Proceedings, Vol. 1552, Issue 1; Conference: 9. international temperature symposium, Los Angeles, CA (United States), 19-23 Mar 2012; Other Information: (c) 2013 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
Country of Publication:
United States
Language:
English