skip to main content

Title: Lens transmission measurement for an absolute radiation thermometer

The lens transmission for the National Institute of Metrology of China absolute radiation thermometer is measured by a hybrid method. The results of the lens transmission measurements are 99.002% and 86.792% for filter radiometers with center wavelengths 633 nm and 900 nm, respectively. These results, after correcting for diffraction factors and the size-of-source effect when the lens is incorporated within the radiometer, can be used for measurement of thermodynamic temperature. The expanded uncertainty of the lens transmission measurement system has been evaluated. It is 1.3×10{sup −3} at 633 nm and 900 nm, respectively.
Authors:
; ;  [1]
  1. Division of Thermometry and Materials Evaluation, National Institute of Metrology, Beijing (China)
Publication Date:
OSTI Identifier:
22218034
Resource Type:
Journal Article
Resource Relation:
Journal Name: AIP Conference Proceedings; Journal Volume: 1552; Journal Issue: 1; Conference: 9. international temperature symposium, Los Angeles, CA (United States), 19-23 Mar 2012; Other Information: (c) 2013 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; DIFFRACTION; FILTERS; LENSES; RADIOMETERS; TEMPERATURE MEASUREMENT; THERMOMETERS; TRANSMISSION; WAVELENGTHS