Title: Lens transmission measurement for an absolute radiation thermometer

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4821399· OSTI ID:22218034
; ;  [1]
  1. Division of Thermometry and Materials Evaluation, National Institute of Metrology, Beijing (China)

The lens transmission for the National Institute of Metrology of China absolute radiation thermometer is measured by a hybrid method. The results of the lens transmission measurements are 99.002% and 86.792% for filter radiometers with center wavelengths 633 nm and 900 nm, respectively. These results, after correcting for diffraction factors and the size-of-source effect when the lens is incorporated within the radiometer, can be used for measurement of thermodynamic temperature. The expanded uncertainty of the lens transmission measurement system has been evaluated. It is 1.3×10{sup −3} at 633 nm and 900 nm, respectively.

OSTI ID:
22218034
Journal Information:
AIP Conference Proceedings, Vol. 1552, Issue 1; Conference: 9. international temperature symposium, Los Angeles, CA (United States), 19-23 Mar 2012; Other Information: (c) 2013 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
Country of Publication:
United States
Language:
English