skip to main content

Title: Assessment of tungsten/rhenium thermocouples with metal-carbon eutectic fixed points up to 1500°C

Four Type A thermocouples and two Type C thermocouples were calibrated at the Au fixed point and Co-C and Pd-C eutectic fixed points. The thermocouples were exposed to 1330 °C for a total of 100 hours. The maximum drift due to the exposure was found to be 4.8 °C. The fixed-point calibration EMF of these thermocouples deviated by less than 0.86% from the temperature specified by the standards ASTM E230-2003 for Type C and GOSTR 8.585-2001 for Type A. The length of one of Type A thermocouples A52 is longer than the others by 150mm. Making use of this provision it was possible to place annealed part of A52 to the temperature gradient part of calibration arrangement every time. Therefore observed aging effect was as low as 0.5 °C compared to the other thermocouples.
Authors:
 [1]
  1. National Research Council of Canada 1200 Montreal Road, Ottawa ON. Canada K1A 0R6 (Canada)
Publication Date:
OSTI Identifier:
22218028
Resource Type:
Journal Article
Resource Relation:
Journal Name: AIP Conference Proceedings; Journal Volume: 1552; Journal Issue: 1; Conference: 9. international temperature symposium, Los Angeles, CA (United States), 19-23 Mar 2012; Other Information: (c) 2013 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
75 CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY; 46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; ANNEALING; CALIBRATION; CARBON; COMPARATIVE EVALUATIONS; ELECTROMOTIVE FORCE; EUTECTICS; RHENIUM; TEMPERATURE GRADIENTS; THERMOCOUPLES; TUNGSTEN