skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Incompatibility of Contrast Medium and Trisodium Citrate

Journal Article · · Cardiovascular and Interventional Radiology
;  [1]
  1. CHU de Charleroi, Department of Radiology (Belgium)

To test the compatibility of trisodium citrate, a catheter lock solution, with iodinated contrast medium. Iohexol, iobitridol, iodixanol, ioxaglate, ioxithalamate, iomeprol, and iopromide were tested. In all tests, 2 ml of contrast medium were mixed with 2 ml of trisodium citrate solution. Iodixanol and ioxaglate provoked a highly viscous gluelike precipitation when mixed with trisodium citrate. A brief transient precipitate was observed with iohexol, iomeprol, and ioxithalamate. Permanent precipitation occurred with iobitridol and iopromide. One must be aware of the potential for precipitation when contrast medium is mixed with trisodium citrate solution. Before trisodium citrate solution is injected, the catheter should be thoroughly flushed with saline if a contrast medium has previously been injected through it.

OSTI ID:
22156500
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 36, Issue 1; Other Information: Copyright (c) 2013 Springer Science+Business Media New York and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); Article Copyright (c) 2012 Springer Science+Business Media, LLC and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English