Title: Weighted simultaneous algebraic reconstruction technique for tomosynthesis imaging of objects with high-attenuation features

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4789592· OSTI ID:22130541
 [1]; ;  [2]; ; ;  [3]
  1. Institute of Medical Engineering, University of Luebeck, Luebeck 23562, Germany and Graduate School for Computing in Medicine and Life Sciences, Luebeck 23562 (Germany)
  2. Institute of Medical Engineering, University of Luebeck, Luebeck 23562 (Germany)
  3. Clinic for Radiology, University Clinics Schleswig-Holstein, Luebeck 23562 (Germany)

Purpose: This paper introduces a nonlinear weighting scheme into the backprojection operation within the simultaneous algebraic reconstruction technique (SART). It is designed for tomosynthesis imaging of objects with high-attenuation features in order to reduce limited angle artifacts. Methods: The algorithm estimates which projections potentially produce artifacts in a voxel. The contribution of those projections into the updating term is reduced. In order to identify those projections automatically, a four-dimensional backprojected space representation is used. Weighting coefficients are calculated based on a dissimilarity measure, evaluated in this space. For each combination of an angular view direction and a voxel position an individual weighting coefficient for the updating term is calculated. Results: The feasibility of the proposed approach is shown based on reconstructions of the following real three-dimensional tomosynthesis datasets: a mammography quality phantom, an apple with metal needles, a dried finger bone in water, and a human hand. Datasets have been acquired with a Siemens Mammomat Inspiration tomosynthesis device and reconstructed using SART with and without suggested weighting. Out-of-focus artifacts are described using line profiles and measured using standard deviation (STD) in the plane and below the plane which contains artifact-causing features. Artifacts distribution in axial direction is measured using an artifact spread function (ASF). The volumes reconstructed with the weighting scheme demonstrate the reduction of out-of-focus artifacts, lower STD (meaning reduction of artifacts), and narrower ASF compared to nonweighted SART reconstruction. It is achieved successfully for different kinds of structures: point-like structures such as phantom features, long structures such as metal needles, and fine structures such as trabecular bone structures. Conclusions: Results indicate the feasibility of the proposed algorithm to reduce typical tomosynthesis artifacts produced by high-attenuation features. The proposed algorithm assigns weighting coefficients automatically and no segmentation or tissue-classification steps are required. The algorithm can be included into various iterative reconstruction algorithms with an additive updating strategy. It can also be extended to computed tomography case with the complete set of angular data.

OSTI ID:
22130541
Journal Information:
Medical Physics, Vol. 40, Issue 3; Other Information: (c) 2013 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English