Title: Feasibility Study of Needle Placement in Percutaneous Vertebroplasty: Cone-Beam Computed Tomography Guidance Versus Conventional Fluoroscopy

ObjectiveTo investigate the accuracy, procedure time, fluoroscopy time, and dose area product (DAP) of needle placement during percutaneous vertebroplasty (PVP) using cone-beam computed tomography (CBCT) guidance versus fluoroscopy.Materials and MethodsOn 4 spine phantoms with 11 vertebrae (Th7-L5), 4 interventional radiologists (2 experienced with CBCT guidance and two inexperienced) punctured all vertebrae in a bipedicular fashion. Each side was randomization to either CBCT guidance or fluoroscopy. CBCT guidance is a sophisticated needle guidance technique using CBCT, navigation software, and real-time fluoroscopy. The placement of the needle had to be to a specific target point. After the procedure, CBCT was performed to determine the accuracy, procedure time, fluoroscopy time, and DAP. Analysis of the difference between methods and experience level was performed.ResultsMean accuracy using CBCT guidance (2.61 mm) was significantly better compared with fluoroscopy (5.86 mm) (p < 0.0001). Procedure time was in favor of fluoroscopy (7.39 vs. 10.13 min; p = 0.001). Fluoroscopy time during CBCT guidance was lower, but this difference is not significant (71.3 vs. 95.8 s; p = 0.056). DAP values for CBCT guidance and fluoroscopy were 514 and 174 mGy cm{sup 2}, respectively (p < 0.0001). There was a significant difference in favor of experienced CBCT guidancemore » users regarding accuracy for both methods, procedure time of CBCT guidance, and added DAP values for fluoroscopy.ConclusionCBCT guidance allows users to perform PVP more accurately at the cost of higher patient dose and longer procedure time. Because procedural complications (e.g., cement leakage) are related to the accuracy of the needle placement, improvements in accuracy are clinically relevant. Training in CBCT guidance is essential to achieve greater accuracy and decrease procedure time/dose values.« less
Authors:
 [1] ; ;  [2] ; ; ; ;  [1]
  1. St. Antonius Hospital, Department of Radiology (Netherlands)
  2. Philips Medical, Department of Clinical Development (Netherlands)
Publication Date:
OSTI Identifier:
22121720
Resource Type:
Journal Article
Resource Relation:
Journal Name: Cardiovascular and Interventional Radiology; Journal Volume: 36; Journal Issue: 4; Other Information: Copyright (c) 2013 Springer Science+Business Media New York and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; ACCURACY; COMPARATIVE EVALUATIONS; COMPUTER CODES; COMPUTERIZED TOMOGRAPHY; CONES; FEASIBILITY STUDIES; FLUOROSCOPY; PATIENTS; PHANTOMS; RADIATION DOSES; TRAINING; VERTEBRAE