Title: Thomson spectrometer-microchannel plate assembly calibration for MeV-range positive and negative ions, and neutral atoms

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4803670· OSTI ID:22118553
 [1]; ; ; ;  [2]; ;  [1];  [3]
  1. Centre for Plasma Physics, School of Mathematics and Physics, The Queen's University of Belfast, Belfast BT7 1NN (United Kingdom)
  2. Max Born Institute for Nonlinear Optics and Short Pulse Spectroscopy, 12489 Berlin (Germany)
  3. WCU Department of Nanobio Materials and Electronics, GIST, Gwangju 500-712 (Korea, Republic of)

We report on the absolute calibration of a microchannel plate (MCP) detector, used in conjunction with a Thomson parabola spectrometer. The calibration delivers the relation between a registered count numbers in the CCD camera (on which the MCP phosphor screen is imaged) and the number of ions incident on MCP. The particle response of the MCP is evaluated for positive, negative, and neutral particles at energies below 1 MeV. As the response of MCP depends on the energy and the species of the ions, the calibration is fundamental for the correct interpretation of the experimental results. The calibration method and arrangement exploits the unique emission symmetry of a specific source of fast ions and atoms driven by a high power laser.

OSTI ID:
22118553
Journal Information:
Review of Scientific Instruments, Vol. 84, Issue 5; Other Information: (c) 2013 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English