Title: Suitability of Exoseal Vascular Closure Device for Antegrade Femoral Artery Puncture Site Closure

Journal Article · · Cardiovascular and Interventional Radiology
; ;  [1];  [2];  [1]
  1. Klinikum Ingolstadt, Department of Diagnostic and Interventional Radiology (Germany)
  2. Klinikum Ingolstadt, Department of Vascular Surgery (Germany)

Purpose. To assess the efficacy and safety of the Exoseal vascular closure device for antegrade puncture of the femoral artery. Methods. In a prospective study from February 2011 to January 2012, a total of 93 consecutive patients received a total of 100 interventional procedures via an antegrade puncture of the femoral artery. An Exoseal vascular closure device (6F) was used for closure in all cases. Puncture technique, duration of manual compression, and use of compression bandages were documented. All patients were monitored by vascular ultrasound and color-coded duplex sonography of their respective femoral artery puncture site within 12 to 36 h after angiography to check for vascular complications. Results. In 100 antegrade interventional procedures, the Exoseal vascular closure device was applied successfully for closure of the femoral artery puncture site in 96 cases (96 of 100, 96.0 %). The vascular closure device could not be deployed in one case as a result of kinking of the vascular sheath introducer and in three cases because the bioabsorbable plug was not properly delivered to the extravascular space adjacent to the arterial puncture site, but instead fully removed with the delivery system (4.0 %). Twelve to 36 h after the procedure, vascular ultrasound revealed no complications at the femoral artery puncture site in 93 cases (93.0 %). Minor vascular complications were found in seven cases (7.0 %), with four cases (4.0 %) of pseudoaneurysm and three cases (3.0 %) of significant late bleeding, none of which required surgery. Conclusion. The Exoseal vascular closure device was safely used for antegrade puncture of the femoral artery, with a high rate of procedural success (96.0 %), a low rate of minor vascular complications (7.0 %), and no major adverse events.

OSTI ID:
22113646
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 36, Issue 3; Other Information: Copyright (c) 2013 Springer Science+Business Media New York and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

The Use of ExoSeal Vascular Closure Device for Direct Antegrade Superficial Femoral Artery Puncture Site Hemostasis
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Cardiovascular and Interventional Radiology · OSTI ID:22113646

Efficacy and Safety of a Novel Vascular Closure Device (Glubran 2 Seal) After Diagnostic and Interventional Angiography in Patients with Peripheral Arterial Occlusive Disease
Journal Article · Mon Apr 15 00:00:00 EDT 2013 · Cardiovascular and Interventional Radiology · OSTI ID:22113646

Efficacy and Safety of the Angioseal Vascular Closure Device Post Antegrade Puncture
Journal Article · Thu May 15 00:00:00 EDT 2008 · Cardiovascular and Interventional Radiology · OSTI ID:22113646