Title: Investigation of a {sup 90}Sr/{sup 90}Y source for intra-ocular treatment of wet age-related macular degeneration

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3213515· OSTI ID:22102096
; ;  [1]
  1. University of Wisconsin Medical Radiation Research Center, Madison, Wisconsin 53705 (United States)

Purpose: The purpose of this study is to perform an extensive investigation of an approximately 2.5 mm long {sup 90}Sr/{sup 90}Y source designed for treating wet age-related macular degeneration. Methods: As part of this investigation, a NIST-traceable absorbed dose to water calibration technique was established, and a source deployment verification test was developed. The influence of treatment cannula construction tolerance on the measurements as well as the dose delivered to the patient was investigated using the Monte Carlo code MCNP5. Variation between production cannulae was quantified experimentally using a well-type ionization chamber, and additional measurements along with Monte Carlo calculations of the collimating insert used for source deployment verification were performed to validate the model. Results: Maximum variation in the integrated target dose was seen when the source was shifted laterally within the treatment cannula. For the well chamber measurements, the observed standard deviation in ionization current for a single source placed in different reference cannulae was {+-}0.3%, with a maximum observed range of less than {+-}0.5%. Clinical cannulae in the collimating insert showed an average of 17.8%{+-}0.4% of the reference signal when sources were fully deployed compared to 18.5% predicted by Monte Carlo calculations. This discrepancy has been attributed primarily to construction of the collimator since the collimation gap was observed to be approximately 0.025-0.075 mm smaller than specified. Construction tolerance of the well chamber insert as well as position tolerance of the cannula tip were both investigated, and their influence on the predicted signal was quantified. Additional measurements along with Monte Carlo based calculations of the collimating insert with polyethylene spacers added to the setup were performed to validate the Monte Carlo model. The shimmed Monte Carlo and measured data agree to within 1%, which is a magnitude difference of approximately 0.1% of the reference signal. Conclusions: This investigation confirms that the signal for an acceptably deployed source in the collimating insert is between 17.5% and 21.5% of the reference signal, as calculated using Monte Carlo models. Clinical cannulae for which the source deployment verification measurement falls outside the acceptable range should not be used to treat patients.

OSTI ID:
22102096
Journal Information:
Medical Physics, Vol. 36, Issue 10; Other Information: (c) 2009 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English