Spatiotemporal Monte Carlo transport methods in x-ray semiconductor detectors: Application to pulse-height spectroscopy in a-Se
- Division of Imaging and Applied Mathematics, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, 10903 New Hampshire Avenue, Silver Spring, Maryland 20993-0002 (United States) and Department of Electrical and Computer Engineering, University of Waterloo, Waterloo, Ontario N2L3G1 (Canada)
Purpose: The authors describe a detailed Monte Carlo (MC) method for the coupled transport of ionizing particles and charge carriers in amorphous selenium (a-Se) semiconductor x-ray detectors, and model the effect of statistical variations on the detected signal. Methods: A detailed transport code was developed for modeling the signal formation process in semiconductor x-ray detectors. The charge transport routines include three-dimensional spatial and temporal models of electron-hole pair transport taking into account recombination and trapping. Many electron-hole pairs are created simultaneously in bursts from energy deposition events. Carrier transport processes include drift due to external field and Coulombic interactions, and diffusion due to Brownian motion. Results: Pulse-height spectra (PHS) have been simulated with different transport conditions for a range of monoenergetic incident x-ray energies and mammography radiation beam qualities. Two methods for calculating Swank factors from simulated PHS are shown, one using the entire PHS distribution, and the other using the photopeak. The latter ignores contributions from Compton scattering and K-fluorescence. Comparisons differ by approximately 2% between experimental measurements and simulations. Conclusions: The a-Se x-ray detector PHS responses simulated in this work include three-dimensional spatial and temporal transport of electron-hole pairs. These PHS were used to calculate the Swank factor and compare it with experimental measurements. The Swank factor was shown to be a function of x-ray energy and applied electric field. Trapping and recombination models are all shown to affect the Swank factor.
- OSTI ID:
- 22098720
- Journal Information:
- Medical Physics, Vol. 39, Issue 1; Other Information: (c) 2012 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
Effect of burst and recombination models for Monte Carlo transport of interacting carriers in a-Se x-ray detectors on Swank noise
The x-ray time of flight method for investigation of ghosting in amorphous selenium-based flat panel medical x-ray imagers
Related Subjects
BIOMEDICAL RADIOGRAPHY
BROWNIAN MOVEMENT
CARRIERS
CHARGE CARRIERS
CHARGE TRANSPORT
COMPTON EFFECT
DIFFUSION
DISTRIBUTION
FLUORESCENCE
MAMMARY GLANDS
MONTE CARLO METHOD
RECOMBINATION
SELENIUM
SEMICONDUCTOR DETECTORS
SIMULATION
SPECTRA
SPECTROSCOPY
THREE-DIMENSIONAL CALCULATIONS
X RADIATION
X-RAY DETECTION