Title: Note: Spatial resolution of Fuji BAS-TR and BAS-SR imaging plates

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4739771· OSTI ID:22093710
; ; ;  [1]
  1. Laboratory for Laser Energetics, University of Rochester, 250 East River Rd., Rochester, New York 14623-1299 (United States)

The spatial resolution of two types of imaging plates, Fuji BAS-TR and Fuji BAS-SR, has been measured using a knife-edge x-ray source of 8-keV Cu K{sub {alpha}} radiation. The values for the spatial resolution, defined as the distance between 10% and 90% levels of the edge spread function, are 94 {mu}m and 109 {mu}m, respectively. The resolution values are important for quantitative analysis of x-ray and particle imaging and spectroscopic diagnostics.

OSTI ID:
22093710
Journal Information:
Review of Scientific Instruments, Vol. 83, Issue 8; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language: