Title: The Optional VenaTech{sup Trade-Mark-Sign} Convertible{sup Trade-Mark-Sign} Vena Cava Filter: Experimental Study in Sheep

Journal Article · · Cardiovascular and Interventional Radiology
 [1];  [2];  [3]
  1. University of Poitiers (France)
  2. Institut National de la Recherche Agronomique (I.N.R.A.) and Assistance Publique-Hopitaux de Paris (APHP), Centre of Research in Interventional Imaging - CRII (France)
  3. Hopital Europeen Georges Pompidou (HEGP) APHP, Department of Respiratory Diseases (France)

Purpose: Retrieval of optional caval filters may be impaired by filter tilting, migration, fracture, or embedding in the IVC wall. The goal of this experimental study was to evaluate a new optional filter, convertible by unlocking and removing the filter head. Methods: Forty-nine Pre-Alp sheep (average weight, 55 kg) were anesthetized. IVC was catheterized via the right femoral vein (n = 46) or via the internal jugular vein (n = 3) with a 12.9-F sheath. VenaTech{sup Trade-Mark-Sign} Convertible{sup Trade-Mark-Sign} IVC filters were inserted as either permanent filters (n = 14) or as filters to be converted. Conversion was immediately after deployment (n = 19) or delayed after 1, 3, or 6 months (n = 20). Filter delivery, deployment, and conversion with measurement of migration and tilting were evaluated by cavography. Incorporation of the filter's stabilizers and arms in the IVC wall was assessed by gross anatomy. Results: Delivery system insertion, filter release, and immediate conversion were successful in all cases. Delayed conversion was completed in all but one sheep, due to insufficient snare tension. Complimentary balloon-catheter inflation was required in 12 of 20 delayed conversions to achieve filter opening. In all 49 sheep, no thrombosis, migration, or significant tilting occurred. Within 4 weeks of conversion, the filter's stabilizers and arms were incorporated into the IVC wall. Upon removal, the filter head was free of intimal growth. Conclusions: The VenaTech{sup Trade-Mark-Sign} Convertible{sup Trade-Mark-Sign} optional IVC filter was successfully implanted in all sheep with no migration or tilting. Conversion at various dates by filter head removal was feasible in all but one case.

OSTI ID:
22082789
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 35, Issue 5; Other Information: Copyright (c) 2012 Springer Science+Business Media, LLC and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

Early and Late Retrieval of the ALN Removable Vena Cava Filter: Results from a Multicenter Study
Journal Article · Mon Sep 15 00:00:00 EDT 2008 · Cardiovascular and Interventional Radiology · OSTI ID:22082789

A Preclinical Study of the Safety and Efficacy of Occlusin Trade-Mark-Sign 500 Artificial Embolization Device in Sheep
Journal Article · Fri Jun 15 00:00:00 EDT 2012 · Cardiovascular and Interventional Radiology · OSTI ID:22082789

'Recovery{sup TM}' Vena Cava Filter: Experience in 96 Patients