Title: Entrance surface dose in cerebral interventional radiology procedures

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4764603· OSTI ID:22075691
; ; ;  [1]
  1. Instituto de Ciencias Nucleares, Universidad Nacional Autonoma de Mexico, 04510 DF (Mexico)

At the Instituto Nacional de Neurologia y Neurocirugia (INNN) diagnostic as well as therapeutic procedures of interventional radiology are carried out. Since the procedures can last from some minutes to several hours, the absorbed dose for the patient could increase dangerously. An investigation had begun in order to determine the entrance surface dose (ESD) using 25 thermoluminiscent dosimeters TLD-100 and 8 strips of 15 Multiplication-Sign 1 cm{sup 2} of Gafchromic XR-QA2 film bound in a holder of 15 Multiplication-Sign 15 cm{sup 2} in the posteroanterior (PA) and lateral (LAT) positions during all the procedure. The results show that maximum ESD could be from 0.9 to 2.9 Gy for the PA position and between 1.6 and 2.5 Gy for the lateral position. The average ESD was between 0.7 and 1.3 Gy for the PA position, and from 0.44 to 1.1 Gy for the lateral position in a therapeutic procedure.

OSTI ID:
22075691
Journal Information:
AIP Conference Proceedings, Vol. 1494, Issue 1; Conference: 12. Mexican symposium on medical physics, Oaxaca (Mexico), 16-19 Mar 2012; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X