Title: Simultaneous translational temperature measurements of different atomic species in plasma flows using scanning Fabry-Perot interferometry

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4719919· OSTI ID:22072298
;  [1]
  1. Institut fuer Raumfahrtsysteme, Pfaffenwaldring 31, 70569 Stuttgart (Germany)

A revised scientific instrument to measure simultaneously kinetic temperatures of different atoms from their optical emission profile is reported. Emission lines are simultaneously detected using one single scanning Fabry-Perot-interferometer (FPI) for a combined spectroscopic setup to acquire different emission lines simultaneously. The setup consists in a commercial Czerny-Turner spectrometer configuration which is combined with a scanning Fabry-Perot interferometer. The fast image acquisition mode of an intensified charge coupled device camera allows the detection of a wavelength interval of interest continuously while acquiring the highly resolved line during the scan of the FPI ramp. Results using this new setup are presented for the simultaneous detection of atomic nitrogen and oxygen in a high enthalpy air plasma flow as used for atmospheric re-entry research and their respective kinetic temperatures derived from the measured line profiles. The paper presents the experimental setup, the calibration procedure, and an exemplary result. The determined temperatures are different, a finding that has been published so far as due to a drawback of the experimental setup of sequential measurements, and which has now to be investigated in more detail.

OSTI ID:
22072298
Journal Information:
Review of Scientific Instruments, Vol. 83, Issue 5; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English