skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Evaluation of the Faraday angle by numerical methods and comparison with the Tore Supra and JET polarimeter electronics

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.3570621· OSTI ID:22062297
; ;  [1];  [2]
  1. CEA, IRFM, F-13108 Saint-Paul-lez-Durance (France)
  2. EURATOM-CCFE Fusion Association, Culham Science Centre, Abingdon, Oxon OX14 3DB (United Kingdom)

On the Tore Supra tokamak, a far infrared polarimeter diagnostic has been routinely used for diagnosing the current density by measuring the Faraday rotation angle. A high precision of measurement is needed to correctly reconstruct the current profile. To reach this precision, electronics used to compute the phase and the amplitude of the detected signals must have a good resilience to the noise in the measurement. In this article, the analogue card's response to the noise coming from the detectors and their impact on the Faraday angle measurements are analyzed, and we present numerical methods to calculate the phase and the amplitude. These validations have been done using real signals acquired by Tore Supra and JET experiments. These methods have been developed to be used in real-time in the future numerical cards that will replace the Tore Supra present analogue ones.

OSTI ID:
22062297
Journal Information:
Review of Scientific Instruments, Vol. 82, Issue 4; Other Information: (c) 2011 EURATOM; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English