Impact of Different Vein Catheter Sizes for Mechanical Power Injection in CT: In Vitro Evaluation with Use of a Circulation Phantom
- University Hospital (RWTH) Aachen, Department of Diagnostic Radiology (Germany)
- University Hospital (RWTH) Aachen, Institute of Medical Statistics (Germany)
The purpose of this study was to evaluate the influence of different peripheral vein catheter sizes on the injection pressure, flow rate, injection duration, and intravascular contrast enhancement. A flow phantom with a low-pressure venous compartment and a high-pressure arterial compartment simulating physiological circulation parameters was used. High-iodine-concentration contrast medium (370 mg iodine/ml; Ultravist 370) was administered in the venous compartment through peripheral vein catheters of different sizes (14, 16, 18, 20, 22, and 24 G) using a double-head power injector with a pressure limit of 325 psi. The flow rate was set to 5 ml/s, with a total iodine load of 36 g for all protocols. Serial CT scans at the level of the pulmonary artery and the ascending and the descending aorta replica were obtained. The true injection flow rate, injection pressure, injection duration, true contrast material volume, and pressure in the phantom during and after injection were continuously monitored. Time enhancement curves were computed and both pulmonary and aortic peak time and peak enhancement were determined. Using peripheral vein catheters with sizes of 14-20 G, flow rates of approximately 5 ml/s were obtained. During injection through a 22-G catheter the pressure limit was reached and the flow rate was decreased, with a consecutive decreased pulmonary and aortic contrast enhancement compared to the 14- to 20-G catheters. Injection through a 24-G peripheral vein catheter was not possible because of disconnection of the canula due to the high flow rate and pressure. In summary, intravenous catheters with sizes of 14-20 G are suitable for CT angiography using an injection protocol with a high flow rate and a high-iodine-concentration contrast medium.
- OSTI ID:
- 21426328
- Journal Information:
- Cardiovascular and Interventional Radiology, Vol. 32, Issue 1; Conference: BSIR 2008: Annual meeting of the Cardiovascular and Interventional Radiology Society of Europe, Manchester (United Kingdom), 5-7 Nov 2008; Other Information: DOI: 10.1007/s00270-008-9359-8; Copyright (c) 2009 Springer Science+Business Media, LLC; ISSN 0174-1551
- Country of Publication:
- United States
- Language:
- English
Similar Records
Effect of venous injection site on accuracy of fast computed tomography (CT) estimation of myocardial perfusion
TH-AB-207A-11: Anatomical and Functional Assessment of Coronary Artery Disease Using Low-Dose Whole-Organ Dynamic Computed Tomography
Related Subjects
BIOMEDICAL RADIOGRAPHY
CAT SCANNING
CONTRAST MEDIA
EVALUATION
FLOW RATE
IN VITRO
INJECTION
IODINE
PHANTOMS
SIZE
TUBES
VEINS
BLOOD VESSELS
BODY
CARDIOVASCULAR SYSTEM
COMPUTERIZED TOMOGRAPHY
DIAGNOSTIC TECHNIQUES
ELEMENTS
HALOGENS
INTAKE
MEDICINE
MOCKUP
NONMETALS
NUCLEAR MEDICINE
ORGANS
RADIOLOGY
STRUCTURAL MODELS
TOMOGRAPHY