skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Initial Clinical Experience Using the Amplatzer Vascular Plug

Journal Article · · Cardiovascular and Interventional Radiology
; ; ; ;  [1]
  1. Leeds Teaching Hospitals NHS Trust, St. James University Hospital, Department of Vascular and Interventional Radiology (United Kingdom)

Background and purpose. The Amplatzer Vascular Plug (AVP) is a self-expanding nitinol wire mesh vascular embolization device derived from the Amplatz septal occluder. We assessed the results of vascular embolization obtained using the AVP. Methods. A retrospective review was carried out of 23 consecutive cases of vascular embolization using the AVP in a variety of different clinical settings. The AVP was chosen to have a diameter approximately 30-50% greater than the target vessel. The device was delivered via an appropriately sized guide catheter and was released when satisfactorily positioned. Additional embolic agents were used in some cases. Results. All target vessels were successfully occluded with no device malpositioning or malfunction. In 14 (61%) patients the AVP was the sole embolic material. In the remaining patients additional agents were used, particularly in preoperative embolization of highly vascular renal tumors. The AVP does not cause instantaneous thrombosis and in high-flow situations thrombosis typically takes up to 15 min. Conclusion. The AVP is a safe, effective embolization device that provides a useful adjunct to the therapeutic armamentarium. It is particularly suited to the treatment of short high-flow vessels where coil migration and catheter dislodgment might occur. In the majority of cases no additional embolic agents are necessary but it may take up to 15 min for complete thrombosis to occur.

OSTI ID:
21090852
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 30, Issue 4; Other Information: DOI: 10.1007/s00270-007-9044-3; Copyright (c) 2007 Springer Science+Business Media, LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

The AMPLATZER Vascular Plug 4: Preliminary Experience
Journal Article · Sun Aug 15 00:00:00 EDT 2010 · Cardiovascular and Interventional Radiology · OSTI ID:21090852

Early Experience with the Amplatzer Vascular Plug II for Occlusive Purposes in Arteriovenous Hemodialysis Access
Journal Article · Mon Feb 15 00:00:00 EST 2010 · Cardiovascular and Interventional Radiology · OSTI ID:21090852

Utility of the Amplatzer Vascular Plug in Splenic Artery Embolization: A Comparison Study with Conventional Coil Technique
Journal Article · Wed Jun 15 00:00:00 EDT 2011 · Cardiovascular and Interventional Radiology · OSTI ID:21090852