Title: Frequent Embolization in Peripheral Angioplasty: Detection with an Embolism Protection Device (Angio Guard) and Electron Microscopy

Journal Article · · Cardiovascular and Interventional Radiology
;  [1];  [2];  [1];  [2]; ;  [1]
  1. University of Tuebingen, Hoppe-Seyler-Strasse 3, 72076Tuebingen, Department of Diagnostic Radiology (Germany)
  2. University of Tuebingen, Hoppe-Seyler-Strasse 3, 72076 Tuebingen, Clinicfor Thoracic and Cardiovascular Surgery (Germany)

Purpose: To evaluate the deliverability and protection capabilities of an embolism protection filter in angioplasty of peripheral arteries. Methods: The Angioguard emboli capture guidewire system was applied in 11 patients with femoropopliteal lesions (6 stenoses, 3 occlusions, 2 controls).Data on lesion crossing, flow deceleration and macroembolization were recorded. Filter membranes were evaluated with scanning electron microscopy (SEM). Results: System delivery was successful in all patients. Primary lesion crossing was feasible in four of six stenoses; predilatation was required in two of six. Marked flow deceleration was recorded in six patients. Emboli next to the filter were detected in each patient with concentric plaques, but could not reliably be removed with the filter. Downstream macroembolization was also present in all patients with concentric stenoses, but in none with chronic occlusion. None of the patients had clinical signs of ischemia. SEM analysis demonstrated only small particles on control group filters and non-obliterating fibrinous conglomerates on filters used in chronic occlusion. Substantial obliteration was seen on several filters used in stenotic lesions. Conclusion:Microembolization of fibrin aggregates is a common incident in balloon angioplasty of femoropopliteal stenoses. Macroembolization occurred more frequently than previously reported. The use of embolism protection filters aided in the detection but not in the removal of larger emboli.

OSTI ID:
21088373
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 26, Issue 4; Other Information: DOI: 10.1007/s00270-003-2656-3; Copyright (c) 2003 Springer-Verlag New York, Inc.; www.springer-ny.com; Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

Retinal Embolization During Carotid Angioplasty and Stenting: Periprocedural Data and Follow-Up
Journal Article · Sun Aug 15 00:00:00 EDT 2010 · Cardiovascular and Interventional Radiology · OSTI ID:21088373

Recanalization of Acute and Subacute Femoropopliteal Artery Occlusions with the Rotarex Catheter: One Year Follow-up, Single Center Experience
Journal Article · Wed Jun 15 00:00:00 EDT 2005 · Cardiovascular and Interventional Radiology · OSTI ID:21088373

Cutting Balloon Angioplasty (CBA) Versus Conventional Balloon Angioplasty (PTA) in the Pre-Dilatation of Carotid Artery Stenosis: Our Preliminary Experience
Journal Article · Thu Nov 15 00:00:00 EST 2007 · Cardiovascular and Interventional Radiology · OSTI ID:21088373