Title: Implication of CT Table Sag on Geometrical Accuracy During Virtual Simulation

Journal Article · · Medical Dosimetry
; ; ;  [1]
  1. Radiation Physics, University of Texas M. D. Anderson Cancer Center, Houston, TX (United States)

Computed tomography (CT) scanners are used in hospitals worldwide for radiation oncology treatment simulation. It is critical that the process very accurately represents the patient positioning to be used during the administration of radiation therapy to minimize the dose delivery to normal tissue. Unfortunately, this is not always the case. One problem is that some degree of vertical displacement, or sag, occurs when the table is extended from its base when under a clinical weight load, a problem resulting from mechanical limitations of the CT table. In an effort to determine the extent of the problem, we measured and compared the degree of table sag for various CT scanner tables at our institution. A clinically representative weight load was placed on each table, and the amount of table sag was measured for varying degrees of table extension from its base. Results indicated that the amount of table sag varied from approximately 0.7 to 6.6 mm and that the amount of table sag varied not only between tables from different manufacturers but also between tables of the same model from the same manufacturer. Failure to recognize and prevent this problem could lead to incorrectly derived isocenter localization and subsequent patient positioning errors. Treatment site-specific and scanner-based laser offset correction should be implemented for each patient's virtual simulation procedure. In addition, the amount of sag should be measured under a clinically representative weight load upon CT-simulator commissioning.

OSTI ID:
21045977
Journal Information:
Medical Dosimetry, Vol. 32, Issue 4; Other Information: DOI: 10.1016/j.meddos.2007.07.001; PII: S0958-3947(07)00119-7; Copyright (c) 2007 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved; Country of input: International Atomic Energy Agency (IAEA); ISSN 0958-3947
Country of Publication:
United States
Language:
English

Similar Records

TU-E-201-03: Eye Lens Dosimetry in Radiotherapy Using Contact Lens-Shaped Applicator
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:21045977

TU-E-201-00: Eye Lens Dosimetry for Patients and Staff
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:21045977

TU-E-201-01: Methods for Eye Lens Dosimetry and Studies On Lens Opacities with Interventionists
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:21045977