Title: Comparison of localization performance with implanted fiducial markers and cone-beam computed tomography for on-line image-guided radiotherapy of the prostate

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
 [1];  [2];  [2];  [2];  [2];  [3];  [2];  [2];  [2];  [2];  [2]
  1. Radiation Medicine Program, Princess Margaret Hospital, Toronto, ON (Canada) and Department of Radiation Oncology, University of Toronto, Toronto, ON (Canada)
  2. Radiation Medicine Program, Princess Margaret Hospital, Toronto, ON (Canada)
  3. Department of Radiation Oncology, University of Toronto, Toronto, ON (Canada)

Purpose: The aim of this work was to assess the accuracy of kilovoltage (kV) cone-beam computed tomography (CBCT)-based setup corrections as compared with orthogonal megavoltage (MV) portal image-based corrections for patients undergoing external-beam radiotherapy of the prostate. Methods and Materials: Daily cone-beam CT volumetric images were acquired after setup for patients with three intraprostatic fiducial markers. The estimated couch shifts were compared retrospectively to patient adjustments based on two orthogonal MV portal images (the current clinical standard of care in our institution). The CBCT soft-tissue based shifts were also estimated by digitally removing the gold markers in each projection to suppress the artifacts in the reconstructed volumes. A total of 256 volumetric images for 15 patients were analyzed. Results: The Pearson coefficient of correlation for the patient position shifts using fiducial markers in MV vs. kV was (R{sup 2} = 0.95, 0.84, 0.81) in the left-right (LR), anterior-posterior (AP), and superior-inferior (SI) directions, respectively. The correlation using soft-tissue matching was as follows: R{sup 2} = 0.90, 0.49, 0.51 in the LR, AP and SI directions. A Bland-Altman analysis showed no significant trends in the data. The percentage of shifts within a {+-}3-mm tolerance (the clinical action level) was 99.7%, 95.5%, 91.3% for fiducial marker matching and 99.5%, 70.3%, 78.4% for soft-tissue matching. Conclusions: Cone-beam CT is an accurate and precise tool for image guidance. It provides an equivalent means of patient setup correction for prostate patients with implanted gold fiducial markers. Use of the additional information provided by the visualization of soft-tissue structures is an active area of research.

OSTI ID:
20944749
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Vol. 67, Issue 3; Other Information: DOI: 10.1016/j.ijrobp.2006.10.039; PII: S0360-3016(06)03363-3; Copyright (c) 2007 Elsevier Science B.V., Amsterdam, Netherlands, All rights reserved; Country of input: International Atomic Energy Agency (IAEA); ISSN 0360-3016
Country of Publication:
United States
Language:
English