Title: Determination of the Sensibility Factors for TLD-100 Powder on the Energy of X-Ray of 50, 250 kVp; 192Ir, 137Cs and 60Co

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.2356420· OSTI ID:20888194
 [1];  [2]
  1. Fac. de Medicina, Universidad Autonoma del Estado de Mexico (Mexico)
  2. SSDL, Departamento de Metrologia ININ, Salazar, Estado de Mexico 15245 (Mexico)

TLD-100 powder is calibrated in terms of absorbed dose to water Dw, using the protocols AAPM TG61, AAPM TG43 and IAEA-TRS 398, for the energy of RX 50, 250 kVp, 137Cs and 60Co respectively. The calibration curves, TLD Response R versus Dw, are fitted by weighted least square by a quadratic polynomials; which are validated with the lack of fit and the Anderson-Darling normality test. The slope of these curves corresponds to the sensibility factor: Fs R/DW, [Fs] = nC Gy-1. The expanded uncertainties U's for these factors are obtained from the ANOVA tables. Later, the Fs' values are interpolated using the effective energy hvefec for the 192Ir. The SSDL sent a set of capsules with powder TLD-100 for two Hospitals. These irradiated them a nominal dose of Dw = 2 Gy. The results determined at SSDL are: for the Hospital A the Dw is overestimated in order to 4.8% and the Hospital B underestimates it in the range from -1.4% to -17.5%.

OSTI ID:
20888194
Journal Information:
AIP Conference Proceedings, Vol. 854, Issue 1; Conference: 9. Mexican symposium on medical physics, Guadalajara, Jal. (Mexico), 18-21 Mar 2006; Other Information: DOI: 10.1063/1.2356420; (c) 2006 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
Country of Publication:
United States
Language:
English