skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Updated Solid Water trade mark sign to water conversion factors for {sup 125}I and {sup 103}Pd brachytherapy sources

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2357018· OSTI ID:20853685
; ; ;  [1]
  1. Department of Radiation Medicine, University of Kentucky, Lexington, Kentucky 40536 (United States)

Dosimetric characteristics of brachytherapy sources are normally determined in water using a Monte Carlo simulation technique and in water equivalent phantom material using both experimental and Monte Carlo simulation techniques. The consensuses of these results are then calculated for clinical applications by converting experimental data obtained in water equivalent material to water using a conversion factor. These conversion factors are normally determined as a ratio of the Monte Carlo-simulated dose rate constant in liquid water to the dose rate constant in a water-equivalent phantom material. However, it has been noted that conversion factors utilized by some investigators have been derived using incorrect phantom material composition and incorrect cross-sectional data information. The impact of errors associated with the cross-sectional data and chemical composition of the phantom material used in dosimetric evaluation of brachytherapy sources has been investigated in this project. Results of these investigations have shown that the use of Solid Water trade mark sign with 1.7% calcium content, as compared to the 2.3% value stated by the manufacturer, may lead to 5% and 9% differences in conversion factors for {sup 125}I and {sup 103}Pd, respectively.

OSTI ID:
20853685
Journal Information:
Medical Physics, Vol. 33, Issue 11; Other Information: DOI: 10.1118/1.2357018; (c) 2006 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English