skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Erbium-doped gold sensor films for magnetic microcalorimeter x-ray detectors

Journal Article · · Journal of Applied Physics
DOI:https://doi.org/10.1063/1.2159207· OSTI ID:20788100
; ; ; ;  [1]
  1. National Institute of Standards and Technology, 325 Broadway MC 817.03, Boulder, Colorado 80305 (United States)

We briefly review the principles of a magnetic microcalorimeter x-ray detector and the requirements for fabricating an array of such sensors, which include a paramagnetic sensor fabricated with thin-film techniques. We discuss two methods for depositing an Er-doped Au sensor film and present magnetization measurements from 4 to 300 K in 5 T and near the detector's operating temperature and field. The properties of the deposited films that we have studied to date deviate from the expected behavior of bulk materials.

OSTI ID:
20788100
Journal Information:
Journal of Applied Physics, Vol. 99, Issue 8; Other Information: DOI: 10.1063/1.2159207; (c) 2006 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0021-8979
Country of Publication:
United States
Language:
English